Parallel Imported One-Alpha Drops x 10 ml – Pharmacy and wholesaler level recall of two batches
Yes – this has been recalled
- The problem
- Pharmacy and wholesaler level recall of two batches of Parallel Imported One-Alpha Drops x 10 ml because the dropper included in this pack delivers 0.05 micrograms of alfacalcidol instead of 0.1 micrograms as labelled on the pack.
- What to do
- Check the batch number on your pack against the list below. Do not stop taking a prescribed medicine without talking to your pharmacist, GP or prescriber first – they can check your supply and swap it if needed.
Source: Medicines and Healthcare products Regulatory Agency (MHRA)
Official notice ↗
Product information
CLDA number
CLDA (17)A/03
MDR number
102-06/17
Parallel import product license holder
Star Pharmaceuticals Ltd
Repackaged by
Servipharm Ltd
Product description
Parallel Imported One-Alpha® Drops x 10ml (alfacalcidol)
2 micrograms/ml Oral drops
Batch number/expiry
| Batch no | Expiry Date |
|---|---|
| A33881 | 10/2019 |
| A37295 | 12/2019 |
Brief description of the problem
The dropper included in this pack delivers 0.05 micrograms of alfacalcidol instead of 0.1 micrograms as labelled on the pack.
Contact details for enquiries
Tel: 020 8423 3800
Email: recall@drugsrus.co.uk
Information if you have the affected batches in stock
Please refer to the form and instructions provided to you by DRUGSRUS directly.